TRANSFER PLATE REPLACEMENT
Tools Required:
3/8” Drive Ratchet•
7/16” Socket •
7/16” Combination Wrench •
Torque Wrench •
Procedure:
Note
Unless otherwise indicated, refer to Figure 4.1 A on pg. 22 for the following replacement procedure.
Completely raise the bed and apply the brakes. 1.
Depending on the siderail position, place the mattress support sections as follows: 2.
Head Siderail: Completely raise the head section. •
Foot Siderail: Lower completely the head section. Completely raise the thigh section, then lift and fold back the •
foot section toward the head end of the bed.
Unplug the bed power cord from the wall outlet. 3.
Lower the siderail needing repair. 4.
Using the 7/16” socket and 7/16” combination wrench, remove the two locknuts, bushings, washers and bolts (I) 5.
holding the transfer plate (J) to the siderail mechanism.
Note
At assembly, tighten the two locknuts to a 75 in/lb torque using the torque wrench.
Reverse the above steps to install the replacement transfer plate. 6.
Verify the siderail for proper operation before returning the bed to service.7.