www.stryker.com 72-1048E MM FL23SE REV A 15
Return To Table of Contents
Problem/Failure Recommended Action
No power to bed Is the bed power switch at the head end of the bed •
turned on?
Is the power cord connected to the power connector •
and plugged into the wall outlet?
Is the power cord severed? Replace if so. •
Are the two fuses inside the power connector still •
operational (see pg. 51)?
Verify power at the wall outlet.•
No bed up or down motion when:
The removable pendant control (optional) is used. •
The foot board command is used.•
Is the Hi-Lo lockout activated (LED off)? If so, deactivate •
it (LED on).
Is the cable of the removable pendant control (optional) •
properly connected to the bed connector under the
mattress support?
Check points of the “No power to bed” problem •
described above.
No Fowler up or down motion when:
The siderail (optional) or removable pendant control •
(optional) is used.
The foot board command is used.•
Is the Fowler lockout activated (LED off)? If so, deactivate •
it (LED on).
Is the cable of the siderail control panel (optional) or •
of the removable pendant control (optional) properly
connected to the bed connector under the mattress
support?
Check points of the “No power to bed” problem •
described above.
The Fowler does not fully raise. This situation arises when the CPR emergency release
(optional) is used to partially lower the Fowler. The use
of the CPR for this purpose creates a situation where the
Fowler motor is out of sync with the actual position of the
Fowler. To correct the situation:
Lower the Fowler • completely using a CPR handle or
the Fowler electric control to enable the Fowler motor
to reset itself and thus synchronize its course with the
actual position (flatten) of the Fowler.
No Knee Gatch up or down motion when:
The siderail (optional) or removable pendant control •
(optional) is used.
The foot board command is used.•
Is the Knee Gatch lockout activated (LED off)? If so, •
deactivate it (LED on).
Is the cable of the siderail control panel (optional) or •
of the removable pendant control (optional) properly
connected to the bed connector under the mattress
support?
Check points of the “No power to bed” problem •
described above.
Troubleshooting Guide