www.stryker.com 72-1048E MM FL23SE REV A 27
Return To Table of Contents
Tools Required:
1/2” Combination Wrench •
Rubber Mallet •
Procedure:
Note
Unless otherwise indicated, refer to Figure 4.1 B on pg. 26 for the following replacement procedure.
Completely raise the bed and apply the brakes. 1.
Unplug the bed power cord from the wall outlet. 2.
Lower the siderail needing repair. 3.
Using the 1/2” combination wrench, remove the two bolts and washers (A) holding the pivot tube ends to the siderail 4.
mechanism arms. Use a rubber mallet to free the tube ends from the arms.
Slide the tube out of its location.5.
Note
Apply grease on a 4” distance from both ends of the tube.
Reverse the above steps to install the replacement pivot tube. 6.
Verify the siderail movement before returning the bed to service.7.
FULL-LENGTH SIDERAIL PIVOT TUBE REPLACEMENT
Service Information