EasyManua.ls Logo

Stryker FL23SE - Foot Board Control Panel Replacement (Optional)

Stryker FL23SE
193 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
34 72-1048E MM FL23SE REV A www.stryker.com
Return To Table of Contents
FOOT BOARD CONTROL PANEL REPLACEMENT (OPTIONAL)
Tools Required:
No. 2 Phillips Screwdriver •
Double-Sided Tape •
Procedure:
Note
The control panel is made of a membrane and its support. Both parts should be ordered for this replacement procedure.
Refer to the “Quick Reference Replacement Parts List” section on pg. 17.
Completely raise the bed and apply the brakes. 1.
Unplug the bed power cord from the wall outlet. 2.
Remove the foot board and lay it on a workbench. 3.
Using the No. 2 Phillips screwdriver, remove the seven screws (A) holding the support cover to the inner face of the 4.
foot board and remove the cover.
Disconnect the foot board panel cable (B). 5.
Using the No. 2 Phillips screwdriver, remove the last screw (C) holding the control panel to the foot board. Remove 6.
the panel. A small part of the panel cable is stuck to the inner face of the panel with a small double-sided tape. Do
the same at reassembly.
Reverse the above steps to install the replacement foot board panel. 7.
Verify all the controls of the foot board panel for proper operation before returning the bed to service. 8.
A
B
C
D
A
Figure 4.2 B
Service Information

Table of Contents

Related product manuals