Tools Required:
No. 2 Phillips Screwdriver •
No. 1 Phillips Screwdriver •
Small Regular Head Screwdriver •
Procedure:
Note
Fuses used in this bed, powered by the 120V electrical system, are of the fast acting type and their rating is 250V, 10A.
For beds powered by other electrical system, see the “Specifications” section on pg. 8 for the fuse ratings.
For the replacement of a fuse, proceed with steps 8 to 10.
Completely raise the bed and apply the brakes. 1.
Unplug the bed power cord from the wall outlet. 2.
Remove the foot board.3.
Raise the head section,using the emeregency crank.4.
Using the No. 2 Phillips screwdriver, remove the seven screws (Q, Figure 4.3 C, 5. pg. 43) holding the cover to the
head casing. Remove all the cables (B) connected to the power connector (D). Note their location so they will be
connected correctly to the replacement power connector. Remove the cover and lay it on a work bench.
Using the No. 1 Phillips screwdriver, remove the two screws (A) holding the power connector to the frame. 6.
Using the small regular screwdriver, remove the fuse holder. 7.
Remove the fuses from the fuse holder and keep them for the replacement power connector, taking care to check 8.
that they are still in good condition, or replace them if you are only proceeding with the replacement of the fuses.
Reverse the above steps to install the replacement power connector. 9.
Turn the power switch on and check that the bed is powered before returning it to service.10.
POWER CONNECTOR AND FUSE REPLACEMENT
Service Information
Figure 4.3 D