www.stryker.com 72-1048E MM FL23SE REV A 49
Return To Table of Contents
Tools Required:
No. 2 Phillips Screwdriver •
1/2” Combination Wrench•
Procedure:
Note
Unless otherwise indicated, refer to Figure 4.3 C on pg. 43 for the following replacement procedure.
Completely raise the bed and apply the brakes. 1.
Unplug the bed power cord from the wall outlet. 2.
Remove the head board. 3.
Using a No. 2 Phillips screwdriver, remove the seven screws (L) holding the cover to the head casing. 4.
Remove the cables connected to the power connector after having carefully noted their position and remove the 5.
cover.
Disconnect the transformer cables. 6.
Using the 1/2” combination wrench, remove the nut, washer and bolt (M) holding the transformer to the head 7.
casing.
Note
Do not over tighten the nut.
Reverse the above steps to install the replacement transformer. 8.
Test all the bed electric controls for proper operation before returning the bed to service.9.
TOROIDAL TRANSFORMER REPLACEMENT (INTERNATIONAL SERIES MODELS)
Service Information