www.stryker.com 72-1048E MM FL23SE REV A 57
Return To Table of Contents
CPR PNEUMATIC CYLINDER REPLACEMENT (OPTIONAL)
Tools Required:
Two 1/2” Combination Wrenches •
11/16” Combination Wrench•
Procedure:
Note
Unless otherwise indicated, refer to Figure 4.3 F on pg. 54 for the following replacement procedure.
Completely raise the bed and apply the brakes. 1.
Completely raise the fowler section. 2.
Unplug the bed power cord from the wall outlet. 3.
Using two 1/2” combination wrenches, remove the nut (K) from the bolt holding the lower end of the pneumatic 4.
cylinder.
Partially remove the bolt until the lower end of the cylinder is free. Keep the shoulder spacer and sleeve (L). 5.
Disengage the lower end of the cylinder from the sleeve still in place.
Using the 11/16” combination wrench, loosen the locknut (M) and unscrew the cylinder completely from the end 6.
fitting (N). Note the location of the locknut before loosening it so the replacement cylinder will be properly screwed
into the end fitting.
Reverse the above steps to install the replacement cylinder. 7.
Test the CPR emergency release for proper operation before returning the bed to service.8.