LUCAS
®
3 Chest Compression System Service Manual
3328798-002, ©2017 Physio-Control, Inc.
Page 74 of 91
Knock the Spring out with a Hammer and a Mandrel or similar tool. Tap the spring downwards with the
mandrel and hammer until it falls loose.
When the Spring is removed from the Angle Shaft, turn the device over so the press fitted side is facing up.
Using the metal Pin or similar tool and Hammer, push the bottom, non press-fitted Angel Shaft out of
Support Leg. Always knock out the non press fitted Angle Shaft first.
Knock the Spring from
round end out of shafts.
Spring with a round end facing up (non-
press fitted side)
Spring with a flat end facing down (press
fitted side)
Spring