LUCAS
®
3 Chest Compression System Service Manual
3328798-002, ©2017 Physio-Control, Inc.
Page 74 of 91
Knock the Spring out with a Hammer and a Mandrel or similar tool. Tap the spring downwards with the
mandrel and hammer until it falls loose.
When the Spring is removed from the Angle Shaft, turn the device over so the press fitted side is facing up.
Using the metal Pin or similar tool and Hammer, push the bottom, non press-fitted Angel Shaft out of
Support Leg. Always knock out the non press fitted Angle Shaft first.
Knock the Spring from
round end out of shafts.
Spring with a round end facing up (non-
press fitted side)
Spring with a flat end facing down (press
fitted side)
Spring

Table of Contents

Other manuals for Stryker LUCAS 3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Stryker LUCAS 3 and is the answer not in the manual?

Stryker LUCAS 3 Specifications

General IconGeneral
Device TypeChest Compression System
Power SourceRechargeable Lithium-ion Battery
Compression Depth5.3 ± 0.2 cm (2.1 ± 0.1 in)
Battery Life45 minutes of operation
Battery Charge TimeApproximately 4 hours
Operating ModesPause
Patient SizeFits most adult patients
Operating Temperature0°C to 40°C (32°F to 104°F)
Storage Temperature–20 to +60 °C (–4 to 140 °F)
IP RatingIP43
Compression Rate100 compressions per minute

Related product manuals