Installation
External Trigger: 1 M
The maximum input voltage at the B NC, between center conductor and shield is 400 V
peak
(DF 39. 2%), 250 V
RMS
to2MHz
derated to 5 V
RMS
at 500 M Hz.
The maximum transient withstand voltage is ±800 V
peak
.
For steady-state sinosoidal waveforms, derate at 20 dB/decade above 200 kHz to 13 V
peak
at 3 MHz and above.
P6139A Passive Probe
Input Voltage:
400 V
RMS
or 400 V DC; CAT I (2, 500 V
peak
transient)
300 V
RMS
or 300 V DC; CAT II (2,500 V
peak
transient
150 V
RMS
or 150 V DC; CAT III (2,500 V
peak
transient)
For steady-state, sinusoidal waveforms, derate at 20 dB/decade above 2.5 MHz to 50 V
RMS
at 20 MHz and above.
Output Voltage (terminated into 1 M):
40 V
RMS
or 40 V DC; CAT I (2,500 V
peak
impulse)
30 V
RMS
or 30 V DC; CAT I (2 50 V
peak
impulse)
15 V
RMS
or 15 V DC; CAT I (2 50 V
peak
impulse)
Temperature:
Operating: -15 °C to +65 °C ( +5 °F to +149 °F)
Nonoperating: -62 °C to +85 °C ( -80 °F t o +185 °F)
Altitude: 2,000 meters
8 DPO4000 Series U ser Manual

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DPO4000 Series and is the answer not in the manual?

Tektronix DPO4000 Series Specifications

General IconGeneral
BrandTektronix
ModelDPO4000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals