Installation
Humidity:
Operating: High: 40 °C to 50 °C, 10% to 60% RH
Operating: Low: 0 °C to 40 °C, 10 to 90% RH
Non-operating: High: 40 °C to 60 °C, 5 to 60% RH
Non-operating: Low: 0 °C to 40 °C, 5 to 90% RH
Altitude:
Operating: 3,000 m (about 10,000 ft)
Nonoperating Altitude: 12,192 m (40,000 ft)
Random Vibration:
Operating: 0.31 G
RMS
, 5 – 500 Hz, 10 minutes per axis, 3 axes (30 minutes to tal)
Non-operating: 2.46 G
RMS
, 5 – 500 Hz, 10 minutes per axis, 3 axes (30 minutes total)
Pollution Degree: 2, I ndoor use only
Acquisition System: 1 MΩ
The maximum input voltage at the BNC, between center conductor and shield is 400 V
peak
(DF ≤ 39.2%), 250 V
RMS
to 130 kHz
deratedto2.6V
RMS
at 500 MHz.
The maximum transient withstand voltage is ± 800 V
peak
.
For steady-state sinusoidal waveforms, derate at 20 dB/decade above 200 kHz to 13 V
pk
at 3 MHz and above.
Acquisition System: 50Ω
The maximum input voltage at the BNC, between center conductor and shield is 5 V
RMS
, with peaks ≤ ±20 V (DF ≤ 6.25%)
DPO4000 Series User Manual 7