EasyManua.ls Logo

Tettex MIDAS micro 2883 - Extended Noise Reduction

Tettex MIDAS micro 2883
138 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
68
Setup
Once the measurement is complete you will be informed in a popup window which will lead back to the setup
screen.
The measured values are now filled into the fields Stray Capacitance and Stray tan δ (the power factor value is
also evaluated and corrected, but not specially displayed).
12.3.3 Extended Noise Reduction
Firmware Version 1.1.0 and greater
The MIDAS micro 2883 incorportates The Extended Noise Reduction
The Midas micro uses special filter algorithms to reduce the noise and extract the measuring signal. See chapter
6.4 for more information.
The extend noise reduction feature is switched on permanently if the “Always On” option is selected. This
guarantees always high accuracy measurement but the measurement speed is reduced. This is the recommended
selection.
If “Auto” is selected the instrument starts automatically the right algorithm whenever a low signal to noise ratio is
detected. In low noise environment this selection optimizes the speed of the measurement.

Table of Contents