EasyManua.ls Logo

ThermoFisher Scientific QuantStudio 12K Flex OpenArray AccuFill System - Page 46

ThermoFisher Scientific QuantStudio 12K Flex OpenArray AccuFill System
103 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1
2
1
Section of the tip box is full
2
Section of the tip box is empty
2.
Select the first section of the sample plate that will be used to fill the OpenArray
Plate.
Select the first section of the sample plate if multiple OpenArray
Plates are filled during a run. The
software selects the total number of sections that correspond with the total number of OpenArray
Plates.
3.
Remove the foil from the appropriate sections of the sample plate, then click the checkbox to
confirm.
Remove the foil only from the sections of the sample plate that are used to load a single
OpenArray
Plate.
The number of sections that the foil is removed from depends on how the run was set up. For
example, if the run was set up without sample information and for two samples per subarray, the
foil is removed from two sections at a time.
Note: Do not remove the foil from all the sections of the sample plate at once.
4.
Close the instrument door.
5.
Click Start Run.
The run does not begin under any of the following conditions:
The waste bin is not in position
The sample plate is not in position
The OpenArray
Plates are not in position
There are more OpenArray
Plates on the instrument deck than are defined in the experiment
setup
The instrument door is open
The Deck screen is displayed.
See Chapter 8, “Monitor run progress”.
IMPORTANT! Each OpenArray
Plate must be prepared for PCR immediately after it is filled (see
“Remove the OpenArray
Plate from the instrument” on page 58).
Chapter 6 Set up and start the run with the quick run workflow
Verify the run setup and start the run—quick run workflow
6
46
QuantStudio
12K Flex OpenArray
AccuFill
System User Guide

Table of Contents

Related product manuals