3.
Select or deselect the Interactive Test checkbox.
If the checkbox is selected, the test includes a step to manually open and close the door.
If the checkbox is not selected, the test is completed without the door being manually opened and
closed.
4.
Enter a number in the Run 1 of field.
The number specified in this field is the total number of times that the instrument performs the
door lock test sequentially.
5.
Click Start test.
The status of each task in the door lock test is displayed.
If more than one run was specified, the number in the Run 1 of field is updated, for example, Run
2 of, Run 3 of.
6.
(For interactive tests only) Follow the prompts in the Door Lock Test dialog box.
If the test fails, contact Service.
Perform a motion test
The motion test moves the robotic tip handler across the dierent axes.
•
X-axis—Left and right
•
Y-axis—Forward and back
•
Z-axis—Up and down
1.
Click Service4Diagnostics.
2.
Select the Motion Test tab.
3.
Select the axes to test.
•
X Axis
•
Y Axis
•
Z Axis
4.
Enter a number in the Run 1 of field.
The number specified in this field is the total number of times that the instrument performs the
motion test sequentially.
5.
Click Start test.
The status of each task in the motion test is displayed.
If more than one run was specified, the number in the Run 1 of field is updated, for example, Run
2 of, Run 3 of.
If the test fails, contact Service.
Appendix D
Manage the service settings
Perform diagnostics tests
D
QuantStudio
™
12K Flex OpenArray
™
AccuFill
™
System User Guide
79