Cleaning and maintenance of the
OpenArray
AccuFill
Instrument
Cleaning schedule and instructions ..................................................... 63
Spill accidents ....................................................................... 64
Cleaning schedule and instructions
Clean the OpenArray
AccuFill
Instrument after each run.
Clean the outside of the instrument
Clean the outside of the OpenArray
AccuFill
Instrument by wiping it with a clean, dry cloth.
IMPORTANT! Do not use solvents.
Clean the inside of the instrument
For routine cleaning, use DNA ExitusPlus reagent or 80–95% ethanol.
Note: For chemical hazard alerts, see Appendix E, “Safety”.
1.
Remove the waste bin from the instrument deck.
CAUTION! Wear appropriate personal protection equipment while handling the waste bin.
2.
Spray the bin with the cleaning solution and wait 10 minutes.
3.
If using DNA ExitusPlus reagent, rinse the bin with tap water to wash away the solution.
4.
Rinse the bin with ethanol.
5.
Allow the bin to dry for 5 minutes before using it again. Make sure the part is completely dry before
loading the next plate.
A
QuantStudio
12K Flex OpenArray
AccuFill
System User Guide
63