Instrument disconnection during run
If the instrument is disconnected from the computer during a run, the run is stopped. The status bar
displays Disconnected (see Figure 4 on page 21).
Reconnect the instrument to the computer (see “Reconnect the instrument” on page 22).
A self-test must be initiated after the instrument is reconnected.
The run must be set up again and restarted for any plates that were not completed during the run that
was interrupted.
Chapter 8 Monitor run progress
Instrument disconnection during run
8
QuantStudio
12K Flex OpenArray
AccuFill
System User Guide
57