If more than one run was specified, the number in the Run 1 of field is updated, for example, Run
2 of, Run 3 of.
Empty the waste bin after the test is complete.
If the test fails, contact Service.
Perform a plate holder test
The plate holder test determines if the instrument holds and senses the OpenArray
Plate correctly.
1.
Click Service4Diagnostics.
2.
Select the Plate Holder Test tab.
3.
Select or deselect the checkboxes that correspond with the OpenArray
Plate holders to test.
OpenArray Plate 1
OpenArray Plate 2
OpenArray Plate 3
OpenArray Plate 4
4.
Enter a number in the Run 1 of field.
The number specified in this field is the total number of times that the instrument performs the
plate holder test sequentially.
5.
Click Start test.
The status of each task in the plate holder test is displayed.
If more than one run was specified, the number in the Run 1 of field is updated, for example, Run
2 of, Run 3 of.
If the test fails, contact Service.
Perform a loading test
The loading test determines if the instrument performs the tasks associated with loading the
OpenArray
Plate correctly.
The test is set up with a quick run workflow (see Chapter 6, “Set up and start the run with the quick run
workflow”).
A sample plate and a waste bin are required.
An OpenArray
Plate and at least one tip box are optional.
Place the items that are used during the loading test on the deck.
1.
Click Service4Diagnostics.
2.
Select the Loading Test tab.
Appendix D
Manage the service settings
Perform diagnostics tests
D
QuantStudio
12K Flex OpenArray
AccuFill
System User Guide
81