The format of the sample plate file is validated. For information about the required format, see “File
formats” on page 11.
The name of the file is displayed in the Select file field.
6.
Repeat step 5 for each CSV file.
7.
Click Next.
The Map plates pane is displayed.
Proceed to “Add or edit sample names—full run workflow” on page 52.
Add or edit sample names—full run workflow
If needed, navigate to the Map plates pane in the Full Run screen.
See Figure 24 on page 49.
See “Options to view a plate layout” on page 36.
If a sample plate file was imported, the sample names are displayed.
The sample plate layout defined in the sample plate file can be edited.
A sample plate file is in CSV format.
If the sample plate file was not imported, the samples must be added manually.
1.
Add or edit the sample name.
2.
Click Next.
Proceed to “Verify the run setup and start the run—full run workflow” on page 52.
Verify the run setup and start the run—full run workflow
See Figure 25 on page 50.
1.
Click each tip box so that the status on the Verify and start run pane matches the physical tip box
in the instrument.
We recommend starting the run with full tip boxes.
The instrument does not start the run if there are not enough tips on the deck.
Chapter 7 Set up and start a run with the full run workflow
Add or edit sample names—full run workflow
7
52
QuantStudio
12K Flex OpenArray
AccuFill
System User Guide