6.
Click Start Run.
The run does not begin under any of the following conditions:
•
The waste bin is not in position
•
The sample plate is not in position
•
The OpenArray
™
Plates are not in position
•
There are more OpenArray
™
Plates on the instrument deck than are defined in the experiment
setup
•
The instrument door is open
The Deck screen is displayed.
See Chapter 8, “Monitor run progress”.
IMPORTANT! Each OpenArray
™
Plate must be prepared for PCR immediately after it is filled (see
“Remove the OpenArray
™
Plate from the instrument” on page 58).
Chapter 7 Set up and start a run with the full run workflow
Verify the run setup and start the run—full run workflow
7
54