6.
Click Start Run.
The run does not begin under any of the following conditions:
The waste bin is not in position
The sample plate is not in position
The OpenArray
Plates are not in position
There are more OpenArray
Plates on the instrument deck than are defined in the experiment
setup
The instrument door is open
The Deck screen is displayed.
See Chapter 8, “Monitor run progress”.
IMPORTANT! Each OpenArray
Plate must be prepared for PCR immediately after it is filled (see
“Remove the OpenArray
Plate from the instrument” on page 58).
Chapter 7 Set up and start a run with the full run workflow
Verify the run setup and start the run—full run workflow