7-1
7 PARAMETERS SETTING MODE
7PARAMETERS SETTING MODE
7.1 Parameter Setting Options
7.1.1 Parameters for Measurement and Display
Menu Selecting Item Contents
1.Ang. Unit deg / gon / mil
Select degree(360°), gon(400G) or mil (6400M) for the
measuring angle unit to be shown on the display .
2.Min.Angle OFF / ON
Select the minimum display angle reading ON or OFF.
GTS-601/601AF[ OFF:1"/ ON:0.5"](0.5mgon/0.1mgon)
GTS-602/602AF[ OFF:5"/ ON:1"] (1mgon/0.2mgon)
GTS-603/603AF[ OFF:5"/ ON:1"] (1mgon/0.2mgon)
GTS-605/605AF[ OFF:5"/ ON:1"] (1mgon/0.2mgon)
3.Tilt OFF / 1axis / 2axis
Select the tilt sensor option for OFF, (1axis) vertical only or
(2axis) vertical and horizontal.
4.Err. corr. OFF / ON
Select the error correction ON or OFF for collimation and error
adjustment.
Note: Perform this item after complete section 8.4 .
For more information, refer to Section 8.4“Adjustment
of Compensation Systematic Error of Instrument” and
Section 8.5“Showing Constant List and Switch ON/
OFF Compensation Systematic Error of Instrument”
5.V -0 Zenith / Level Select the vertical angle reading for Zenith 0 or Horizontal 0.
6.HAmem OFF / MEM.ON
It is able to retain presetting angle after turning power
off.(MEM.ON )
Note: After changing this parameter, turn the power switch
off once.
7.Light OFF / ON
Select the option to turn on the backlight when the power is
turned on.
Note: Resume Mode ON/OFF option is given priority to
regardless of the setting of this item.
Prog
Std
Mem
Com
Adj
Para
2001-07-10 15:30:40
PARAMETERS SETTING MODE
[Press [F6] key.]
In this mode, setting of parameters regard with measuring, displaying and
communications will be done.
When a parameter is changed and set, the new value is stored into memory.Press [F6]
key from the main menu icons, the following display will be shown.
The parameter modes is classified in Measurement and Communication.
Parameters
F1 Measurement
F2 Communication
F3 Password