48 AU-OPR-AureFloFT-EN,
Rev H
EMC Tables AureFlo System
Table 9 - Test Specications for Enclosure Port Immunity to RF Wireless Communications
Equipment
TEST
FREQUENCY
(MHz)
BAND
(MHz)
SERVICE MODULATION
MAXIMUM
POWER (W)
DISTANCE
(m)
IMMUNITY TEST
LEVEL (V/m)
385 380-390 TETRA 400
Pulse Modulation
18 Hz
1.8 0.3 27
450 430-470
GMRS 460,
FRS 460
FM
± 5 kHz deviation
1 kHz sine
2 0.3 28
710
704-787 LTE Band 13, 17
Pulse Modulation
217 Hz
0.2 0.3 9745
780
810
800-960
GSM 800/900,
TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5
Pulse Modulation
18 Hz
2 0.3 28
870
930
1,720
1,700-
1,990
GSM 1800,
CDMA 1900,
GSM 1900,
DECT, UMTS,
LTE Band 1, 3,
4, 25
Pulse Modulation
217 Hz
2 0.3 28
1,845
1,970
2,450
2,400-
2,570
Bluetooth,
WLAN,
802.11 b/g/n,
RFID 2450,
LTE Band 7
Pulse Modulation
217 Hz
2 0.3 28
5,240
5,10 0 -
5,800
WLAN 802.11
a/n
Pulse Modulation
217 Hz
0.2 0.3 95,500
5,785
Note: The meter may display incorrect ow values, momentarily stop working or need to be turned off and then turned on again (a very rare
occurrence) when in close proximity to high levels of RF wireless communication signals.