User's Manual
750 50 0042 03 70
13.5 Page 5
21. Activation of manual temperature compensation.
When this function is activated, the measured value is
compensated according to the instrument and part
temperatures and their expansion coefficient. The
displayed value is brought to a temperature of 20 °C
(68 °F).
22. Instrument temperature
23. Part temperature
24. Expansion coefficient of the part
When temperature compensation is activated, a symbol appears in the status bar of
the main screen. This function can result in large measurement deviations. It should
therefore be used with the greatest care!
The temperature values must be ebtered in degree Celsuis (°C).
13.6 Page 6
25. Measuring force and carriage balancing:
The measuring force can be adjusted by clicking on
the list and selecting the desired force. To adjust the
balancing of the carriage, refer to § 14.1 .
26. Probing sensitivity and 2nd line of the display:
Probing sensitivity can be adjusted, depending on the
type of insert used :
- SHORT: For short and stiff inserts
- MEDIUM: For medium long
- LONG: For long and less rigid inserts
Note:
Probing filters "MEDIUM" and "LONG" reduce the
probing robustness. Precision and repeatability can
differ from specifications according to the user.
In height measuring mode, the 2nd line of the display
can be hidden. To do this, uncheck the check box.
27. Date of last and next calibration:
The date of the next calibration can be entered
manually. There is no function linked to this date, only
user information.
28. Model, type, serial no. and firmware version