NOTE
If a Model 3440 Plus equipped with the GPS option is
moved a long distance between uses, the GPS system
must be allowed to initialize. In some instances,
initialization may take as long as 30 to 45 minutes from
the time the gauge is powered on with the GPS enabled.
Note also that the gauge must be positioned such that
the GPS receiver can receive signals from the GPS
satellites (see Appendix I for more information). If the
GPS does not initialize within 45 minutes, contact your
Troxler representative.
NOTE
If the gauge display is difficult to read in bright light,
adjust the contrast as described in the Display Contrast
section on page C–9.
After 5 hours of no activity, the gauge automatically performs a
total power shutdown.
NOTE
If the charge calibration (see page C–11) is BAD, the
gauge will not perform an automatic shutdown.