7–4
MOISTURE OFFSET
The moisture offset (k) is determined by comparing the moisture of
a laboratory sample with the gauge moisture reading. The gauge
allows the operator to enter the moisture offset manually, or to have
the gauge derive the moisture offset. The gauge can store up to four
moisture offsets.
To access the Moisture Offset menu, press 2 at the Offset
menu shown on page 7–2. The gauge displays:
The Moisture Offset menu shows the moisture offset values
stored in memory (if any) and the New and Disable options.
ENABLE A STORED MOISTURE OFFSET
To enable a moisture offset value displayed on the Moisture
Offset menu, press the number key that corresponds to that value.
The gauge enables the moisture offset value and returns to the
Ready screen.
ENTER A NEW MOISTURE OFFSET
A new moisture offset may be entered either manually using the
keypad or by using the gauge to record actual measurements, then
entering the true moisture value as determined by an alternate
method. To enter and enable a new moisture offset value, press
5
at the Moisture Offset menu shown above. The gauge displays:


   
   
   



  
  