Model 3430 Plus & 3440 Plus
7–7
As prompted, place the gauge on the measurement site and press
any key. The gauge takes a 240-second count, during which it
displays the progress of the count. Upon completion, the gauge
calculates and displays the moisture offset (K) as shown:
To enable and use this moisture offset without storing it for later
use, press
NO. The gauge enables the offset and returns to the
Ready screen.
To enable the moisture offset and store it for later use, press
YES. The gauge displays:
The gauge can store the value in one of four memory cells.
Storing a new value in a cell erases the old value. Use the
number keys to store the value. This will replace any stored
values that were previously associated with the memory cell.
The gauge stores the moisture offset, enables it for use, and
returns to the Ready screen.
DISABLE THE MOISTURE OFFSET
To disable the moisture offset value, press 6 at the Moisture
Offset menu shown on page
7–4. The gauge disables the moisture
offset and returns to the Ready screen.
NOTE
The mositure offset is also disabled when the gauge is
turned off.
 
  
  



   
   
  
7. CALIBRATION OFFSETS