Model 3450
5–15
PRECISION
The gauge can produce test results with precision limits as low as
0.1 pcf (1.6 kg/m
3
) under certain conditions. The Precision function
is used to determine the count time required to achieve the precision
entered by the operator. If the precision requested is within range
and the time required to calculate the required result does not
exceed 60 minutes, the following procedure can usually produce the
required results.
To execute the Precision function, first ensure that the gauge is in
the correct mode: Soil or Asphalt. Press
7 from the Setup menu
shown on page
5–3. The gauge displays:
Enter the desired precision and press ENTER/START. The gauge
displays:
If the gauge is in the correct position, press ENTER/START. The
gauge displays:

 
  


  
  


 
  
5. SETUP MENU