Model 3450
5–9
REVIEWING THE DRIFT TEST
To review the results of the drift stat test, press 2 at the Drift Test
menu shown on page
5–7. The gauge displays the results of the
density drift test as shown on page 5–8
. Press ENTER/START to
view the results of the moisture drift test. Follow the prompts to
scroll through the results.
PRINTING THE DRIFT TEST
To print the results of the last drift test to a serial printer or
computer via the gauge’s 9-pin serial port, press
3 at the Drift
Test menu shown on page
5–7. The gauge prompts the operator to
connect the printer to the gauge.
Connect a serial cable to the gauge’s serial port. Ensure that the
serial cable meets the pinout shown on page F–6. An optional serial
cable, PN 113128, is available from Troxler. Connect the serial
cable to the printer or computer serial port.
Press ENTER/START. The gauge prints or uploads the drift test
data and returns to the Drift Test menu.