5–6
Moisture Offset – Gauge Derived
From the last display on the previous page, select 1 for the display:
Select:
1- True M x.x %
2-Gauge M x.x %
ENTER to enable
For a true gauge-derived offset, two values must be input: the true
moisture value, which is determined by laboratory analysis, and the gauge
moisture, which is the gauge reading on the material to be tested.
If the above values are correct, press START/ENTER.
To input a new true moisture value, press 1 for the display:
True Moisture-
x.x %
Press ENTER
when completed
Input the laboratory value and press START/ENTER.
To input a new gauge moisture value, press 2 for the display:
Gauge Moisture-
Select:
1- Keypad entry
2- Measurement
KEYPAD ENTRY. To use the keypad to enter the gauge moisture reading
from a previous test, press 1. The display is:
Gauge Moisture-
x.x %
Press ENTER
when completed