25
When using the matching box's panel with punched holes for ventilation and the fan installation
panel as the top panel, there is a risk that contaminants will enter the matching box from that top
panel and short circuit the internal circuits or cause arcing due to insulation deterioration, so take
care to use the maintenance panel as the top panel. If there is no way to prevent using the panel
with punched holes as the top panel due to the structure, take measures to ensure that there is no
risk of contaminants entering the matching box from the top panel by ensuring clearance of 10 mm
or more.
3-3-2. In the case of connector output
On the standard MBX-1305N/MBX-1310N, output is from the base by an output busbar,
but connector output can be selected through an option. The HN model and DIN_7/16
model can be selected for the connector output. However, take care during usage as the
input power and load impedance may cause the output voltage and output current to
exceed the specifications of the connector and cable due to a sympathetic vibration
phenomenon at the matching box output section.
[Application conditions]
Model: MBX-1305N/MBX-1310N
Output current: 10 A or less*
Output voltage: 5,000 Vpp or less*
Output cable length: 1 m or less
*Usage is not possible when this is exceeded due to the input power and load
impedance.