119
1 EXP. HOLD FUNCTIONS
119
EXP. HOLD FUNCTIONS
Chapter
1
1-6. Hold End Timing
The HOLD END (H/E) signal timing when Inflection Point, Relative Maximum, Relative
Minimum or Relative Difference hold is used can be changed.
Hold Sct / Detect Sct Initial value: Detect Sct
How to set
- Hold Sct: HOLD END (H/E) is on according to timing in which the SECTION signal is
off.
- Detect Sct: HOLD END (H/E) is on immediately after the detection of the hold point.
■Hold operations
SET. Exp. Hold Set. Page 2
→→
Exp. Setting
→
In the case of Inflection Point hold
SECTION
T/H
Detection Hold Section
H/E
HOLD Status OFF ON OFF
INF Status OFF Blink ON OFF
When the HOLD END t g is "Detection section"
H/ E
OFF ON OFF
INF Status OFF Blink ON OFF
Tracking Section
Indicated value
Sensor Input Value
imi
n
When the HOLD END timing is "Hold section"
Tracking Section