4 CALIBRATION
37
CALIBRATION
Chapter
4
4-7.
Equivalent Input Calibration
Set the rated output value and reading of the sensor.
Rated output value: -3.0003.000mV/V0 is excluded.
Display value:    -99999999990 is excluded.
How to set
1. Press the [SET.] button.
(refer to page 36)
2. Press the [Calibration] button.
(refer to page 36)
3. Press the [Equiv.Cal.] button.
4. After pressing the [Rate Out] button,
enter the rated output of the sensor
with the numerical keys.
5. After pressing the [Disp Val] button,
enter the display value with the
numerical keys. Also, set the decimal
place here. Press the [OK] button to
perform calibration. To eliminate a
decimal point, register by inputting a
numerical value and then pressing the
decimal point key at the end.
Registration method at decimal place
Example)
The indicated value is made "200.0" by Equivalent Input Calibration.
It is a specification to which the decimal point is input with the numerical keys.
Please input it in the image that operates the calculator.
Please put it again from the start pushing when you correct it.
→→
The display of the decimal point
Press
of the input value moves.
The decimal point ten keys are registered pushing at the end after the
numerical value is input when the decimal point is lost.
All set values concerning the load synchronize with the decimal place of the
indicated value (capacity).