2 EXP. COMPARISON FUNCTIONS
120
EXP. COMPARISON FUNCTIONS
Chapter
2
2-1. Before Value Comparison
Before value comparison
ON / OFF Initial value: OFF
How to set
In addition to the HI/LO Limit comparisons of hold values, the difference between the presently-
measured hold value and the previously-measured hold value can be judged.
Difference value = Present hold value (indicated value) - Previous hold value
In the case of Peak hold
* The comparison output in the above timing chart applies to the case where the Comparison Timing setting in
Comparison Setting is Hold.
* The before value is renewed when the hold value is canceled.
2
EXP. COMPARISON FUNCTIONS
SET. Exp. Comp. Set. Page 1
→→
Exp. Setting
5'%6+10 㪦㪥 㪦㪥 㪦㪥
6* 㪦㪥 㪦㪥
*' 㪦㪥 㪦㪥 㪦㪥
*+ 㪦㪥
.1 㪦㪥
1-
&(A*+
㪦㪥 㪦㪥
&(A.1 㪦㪥
*+
.1
&KH*+
&KH.1
The before value is
renewed when the
hold value is canceled.
Difference from the previous hold value