5 SETTING OF FUNCTIONS
65
SETTING OF FUNCTIONS
Chapter
5
Graph mode
Set the graph plotting mode.
Setting range: Continued, Ext, Level, Ext+Level
How to set
Interval time
If you select "Continued" in the graph function setting, set the graph plotting operation interrupting
time from clearing the screen until moving to the next graph plotting operation. During this time,
the graph screen is held.
Setting range: 00.0 - 99.9 sec.
How to set
Graph start level
If you select [level] or [external + level] in the graph function setting, set the graph plotting start
level.
Setting range:99999 - 99999
How to set
Level detection conditions
If you select [level] or [external + level] in the graph function setting, set the graph plotting start
level comparison conditions.
Setting range: Passed, Passed HI, Passed LO, Beyond, Below
How to set
SET. Graph Setting Page 1
→→
Work Setting
SET. Graph Setting Page 2
→→
Work Setting
SET. Graph Setting Page 1
→→
Work Setting
SET. Graph Setting Page 2
→→
Work Setting
- Passed
Graph plotting starts when the indicated value crosses the level set value.
- Passed HI
Graph plotting starts when the indicated value crosses the level set value upward.
- Passed LO
Graph plotting starts when the indicated value crosses the level set value
downward.
- Beyond
Graph plotting starts when the indicated value is larger than the level set value.
- Below
Graph plotting starts when the indicated value is smaller than the level set value.