41
5 SETTING OF FUNCTIONS
41
SETTING OF FUNCTIONS
Chapter
5
5-2. Digital Filter
The digital filter is a function for reducing drifts of the indicated value by means of a moving
average of data converted from analog to digital. With an increase in the number of filterings, the
indicated value becomes more stable, but the response to inputs becomes slower.
Number of settings: OFF, 2 - 999
How to set
5-3. Analog Filter
A low-pass filter is provided for filtering input signals from the strain-gage sensor and canceling
noise components.
The cut-off frequency can be selected in a range between 30Hz and 1000Hz. With an increase in the
cut-off frequency, the response becomes faster, but noise components may be indicated.
Cut-off frequency: 30Hz, 100Hz, 300Hz, 1000Hz
How to set
5-4. Motion Detect (MD)
Enter the parameters to detect stable.
If the difference between the current indicated value and the 50-msec-previously indicated value is
less than the set count and the duration of the condition is more than the set time, the indicated
value is regarded to be stable.
Setting range:
MD (Time): 0.0 - 9.9 sec.
MD (Count): 0 - 99 count
How to set
SET. Operation Page 1
→→
System Setting
→
SET. Operation Page 1
→→
System Setting
→
Setting
Indicated Value
50msec
Time
Setting Time
Count
SET. Operation Page 2
→→
System Setting
→
When the time is 0.0 sec. and the width is 00 markings, stable is not detected.
Stability detection by Motion Detect is closely related to the SI/F print (stable
value) function and comparison timing. For details, see "5-8.SI/F Print Out"on
page 43 and " ■ Comparison timing"on page 46.