38
4 CALIBRATION
38
CALIBRATION
Chapter
4
4-8. Actual Load Calibration
Set the actual load value under an actual load.
Setting range:  -99999~999990 is excluded.
How to set
1. Press the [SET.] button.
(refer to page 36)
2. Press the [Calibration] button.
(refer to page 36)
3. Press the [Actual Cal.] button.
4. Apply an actual load to the sensor, enter
the actual load value by the numerical
keys and determine with the [OK] button.
Also, set the decimal place here. To
eliminate a decimal point, register by
inputting a numerical value and then
pressing the decimal point key at the end.
4-9. Increment (This step may be omitted if there is no change.)
Set the increment of the indicated value.
1 、 2 、 5 、 10 、 20 、 50 、 100(The display changes by the decimal place.
How to set
1. Press the [SET.] button. (refer to page 36)
2. Press the [Calibration] button. (refer to page 36)
3. Select page 2,
and press the
[Increment]
button.
4. Select the
increment and determine with the
[OK] button.
SET. Calibration Page 1
→→
SET. Calibration Page 2
→→