EasyManua.ls Logo

Unipulse F372A - 4-2. Calibration Protect; 4-3. Calibration Value Selection

Unipulse F372A
156 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
34
4 CALIBRATION
34
CALIBRATION
Chapter
4
4-2. Calibration Protect
Calibration-related set values can be protected so that they will not be changed by misoperation.
When Cal. Protect is ON, no change can be made while the alarm sounds.
ON Protected
OFF Unprotected
How to set
1. Press the [SET.] button.(refer to page 36)
2. Press the [Protect / Init.] button.
3. Press the [Cal. Protect] button.
4. Select the ON/OFF and determine with
the [OK] button.
4-3. Calibration Value Selection
(This step may be omitted if there is no change.)
By storing up to four calibration values in the memory, the desired calibration value can be called to
switch the indicated value. Setting values that can be switched are as follows:
Calibration value selection by touch panel
Select "0" - "3" by pressing the button on the CALIBRATION screen.
When not using, set "0".
SET. Protect / Init. Page 1
→→
Calibration Mode Setting
Excitation Voltage
Zero Calibration
Actual Load Calibration
Equivalent Input Calibration
Increment
Unit
Digital Offset
DZ Limit

Table of Contents

Related product manuals