79
6 OPTION
79
OPTION
Chapter
6
- Taking voltage output signals
Connect external equipment (2k or more
load resistance) to + and - of the F372A.
- Taking current output signals
Connect external equipment (350 or less
load resistance) to CUR and G of the
F372A.
- Resolution
The D/A converter has a resolution of 1/10000 to -10 - +10V (4 - 20mA).
Scale setting value selection
Four types of zero scale and full scale settings can be stored.
Switch with the selector button ( / ) to conduct scale setting.
Scale of output of D/A converter changes accordingly when gain changes by the selection of
calibration value.
How to set
Zero scale output value and full scale output value
Setting range: 99999 - 99999 (where zero set value < full scale set value)
- In the case of voltage output
Zero scale: Set the indicated value for the time of output of 0V.
Full scale: Set the indicated value for the time of output of 10V.
- In the case of current output
Zero scale: Set the indicated value for the time of output of 4mA.
Full scale: Set the indicated value for the time of output of 20mA.
F.G, shield, etc.
Inside Outside
F372A
External equipment
Load resistance:
2k or more
F.G, shield, etc.
F372A
Inside Outside
External equipment
Load resistance:
350 or less
CAUTION
- The D/A converter is optional.
- Do not apply external voltage because it will break.
- Do not short-circuit the voltage output because a failure will occur.
Also, connecting a capacitive load may cause oscillation.
SET. Option Setting (D/A OUTPUT) Page 1
→→