EasyManua.ls Logo

Unipulse F372A - Communication Format

Unipulse F372A
156 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
71
5 SETTING OF FUNCTIONS
71
SETTING OF FUNCTIONS
Chapter
5
Communication format
1.Normal
- Reading the indicated value (sign, 5-digit indicated value, decimal point)
- Reading the status (7-digit)
*The hold bit is operated at the same timing as the H/E signal.
Host
F372A
RA CR
RA
 Delimiter
1 0 0 0 0 .
* If the indicated value has no decimal point, a decimal point follows the numerical value.
Host
F372A
RD CR
R D 0 0 0 0 0 0 0
1:ON
0:OFF
1:ON
0:OFF
Near Zero 1:ON
Output signal 0:OFF
HI 1:ON
Output signal 0:OFF
LO 1:ON
Output signal 0:OFF
OK 1:ON
Output signal 0:OFF
Undefined
*Hold
Stable
 Delimiter
Host
F372A
RC CR
R C 0 0 0 0 0 0 0
1:ON
0:OFF
1:ON
0:OFF
1:ON
0:OFF
1:ON
0:OFF
LL 1:ON
Output signal 0:OFF
HH 1:ON
Output signal 0:OFF
Overload
Sensor +error
Delimiter
Output signal 0:OFF
1:ON
Sensor -error
+OVER
-OVER

Table of Contents

Related product manuals