32
4 CALIBRATION
32
CALIBRATION
Chapter
4
About calibration
Calibration is performed for matching the F372A to a strain-gage sensor. The following two types
of calibration are available for the F372A.
● Equivalent input calibration
Calibration is performed without an actual load by entering the rated output value (mV/V) and
the capacity (to be indicated) of the strain-gage sensor by the keys. Calibration is easily
performed when no actual load is available.
For example, the gain is automatically determined by entering:
2.001mV/V (rated output) - 100.0kg (capacity)
● Actual load calibration
Apply an actual load to the strain-gage sensor and enter the actual load value by the keys for
calibration. Calibration is accurately performed with reductions in errors.
F372A
Strain gauge sensor
Rated output value (mV/V)+Indicated value
A data sheet is attached to a strain-gage sensor at the time of purchase.
Enter the rated capacity value and the rated output value described on
the data sheet into the F372A.
Point
Strain gauge sensor
Indicated value
F372A
Actual
load