128
2 EXP. COMPARISON FUNCTIONS
128
EXP. COMPARISON FUNCTIONS
Chapter
2
2-4. Before Value Comparison and Relative Value Comparison
(in expansion Double Hold only)
In Double Hold, Before Value Comparison and Relative Value Comparison can be made
simultaneously.
Screens appearing when before value comparison &
relative value comparison are selected
Before-value-comparison & Relative-Value-Comparison-specific measurement screens appear.
Also, the Comparison Setting menu is dedicated.
Difference Value- and relative-HI/LO limit comparisons
HI/LO Limit, Difference value-HI/LO, and Relative-HI/LO Limit comparisons of Double Hold are
made, and respective judgment results are output externally.
When the Comparison Standard is A:
The HI/LO Limit comparison result of the relative value, and the HI/LO Limit comparison result
and Difference value-HI/LO Limit comparison result of hold value A are output together.
- Hold value A HI Limit (HI-A) & Difference value-HI Limit (DF-HI)
ON condition: Hold value A > HI-A Limit, or Difference value > Difference value-HI Limit
OFF condition: Hold value A
HI-A Limit, and Difference value
Difference value-HI Limit
- Hold value A LO Limit (LO-A) & Difference value-LO Limit (DF-LO)
ON condition: Hold value A < LO-A Limit, or Difference value < Difference value-LO Limit
OFF condition: Hold value A
LO-A Limit, and Difference value
Difference value-LO Limit
- Relative-HI Limit (RL-HI)
ON condition: Relative Value (Hold value B - Hold value A) > Relative-HI Limit
OFF condition: Relative Value (Hold value B - Hold value A)
Relative-HI Limit
- Relative-LO Limit (RL-LO)
ON condition: Relative Value (Hold value B - Hold value A) < Relative-LO Limit
OFF condition: Relative Value (Hold value B - Hold value A)
Relative-LO Limit
- OK
ON condition: All of HI-A, LO-A, DF-HI, DF-LO, RL-HI, and RL-LO are OFF.
OFF condition: Any of HI-A, LO-A, DF-HI, DF-LO, RL-HI, and RL-LO is ON.
* The Hysteresis setting is not reflected in the Difference-HI/LO Limit and Relative-HI/LO Limit
comparison operation.
* If the Comparison Timing condition is not met, the comparison result of the value targeted for the
Comparison Timing turns OFF.
* OK is displayed each on the LCD screen as hold value A and the difference value are OK, and the
relative value is OK
Rel.
Hold A
Hold B
Dif.
Hold A
Hold B