EasyManua.ls Logo

Unipulse F372A - Load Cell Connection

Unipulse F372A
156 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
23
2 INSTALLATION & CONNECTION
23
INSTALLATION & CONNECTION
Chapter
2
Load cell connection
- 4-wire sensor
- 6-wire sensor
Short-circuit +EXC with +S and -EXC with -S for connecting a 6-wire strain-gage sensor.
Cable colors of sensors
Cable colors of sensors may differ from one manufacturer to another (it may even differ from one
model to another for some products). Refer to the sensor manual (or data sheet) and check signal
names and colors in order to connect the cables correctly.
( EXC)
( EXC)
( SIG)
( SIG)
(SHIELD)
IN
OUT
IN
OUT
1
2
3
4
F372A
5
( EXC)
( EXC)
( SIG)
( SIG)
IN
OUT
IN
OUT
F372A
1
2
3
4
(SHIELD)
5

Table of Contents