2 EXP. COMPARISON FUNCTIONS
129
EXP. COMPARISON FUNCTIONS
Chapter
2
.
When the Comparison Standard is B:
The HI/LO Limit comparison result of the relative value, and the HI/LO Limit comparison result
and Difference value-HI/LO Limit comparison result of hold value B are output together.
- Relative-HI Limit (RL-HI)
ON condition: Relative Value (Hold value A - Hold value B) > Relative-HI Limit
OFF condition: Relative Value (Hold value A - Hold value B)
Relative-HI Limit
- Relative-LO Limit (RL-LO)
ON condition: Relative Value (Hold value A - Hold value B) < Relative-LO Limit
OFF condition: Relative Value (Hold value A - Hold value B)
Relative-LO Limit
- Hold value B HI Limit (HI-B) & Difference value-HI Limit (DF-HI)
ON condition: Hold value B > HI-B Limit, or Difference value > Difference value-HI Limit
OFF condition: Hold value B
HI-B Limit, and Difference value
Difference value-HI Limit
- Hold value B LO Limit (LO-B) & Difference value-LO Limit (DF-LO)
ON condition: Hold value B < LO-B Limit, or Difference value < Difference value-LO Limit
OFF condition: Hold value B
LO-B Limit, and Difference value
Difference value-LO Limit
- OK ON condition: All of HI-B, LO-B, DF-HI, DF-LO, RL-HI, and RL-LO are OFF.
OFF condition: Any of HI-B, LO-B, DF-HI, DF-LO, RL-HI, and RL-LO is ON.
* The Hysteresis setting is not reflected in the Difference-HI/LO Limit and Relative-HI/LO Limit
comparison operations.
* If the Comparison Timing condition is not met, the comparison result of the value targeted for the
Comparison Timing turns OFF.
* OK is displayed each on the LCD screen as the relative value is OK, and hold value B and the
difference value are OK.
A1
COM1 COM2
B1
A2 OUT RL-HI WORK0 IN B2
A3 OUT HI-A & DF-HI WORK1 IN B3
A4 OUT OK WORK2 IN B4
A5 OUT LO-A & DF-LO WORK3(LOCK) IN B5
A6 OUT RL-LO T/H IN B6
A7
COM1 COM2
B7
A8 OUT HOLD END(H/E) SECTION IN B8
A9 OUT RUN D/Z IN B9
A10 OUT EVENT GRAPH TRIG IN B10
A11 OUT SIF CAL0 IN B11
A12 OUT SIF CAL1 IN B12
- External I/O pin assignments for Comparison Standard A
FCN-365P024-AU/FCN-360C024-B
External I/O
A1
COM1 COM2
B1
A2 OUT HI-B & DF-HI WORK0 IN B2
A3 OUT RL-HI WORK1 IN B3
A4 OUT OK WORK2 IN B4
A5 OUT RL-LO WORK3(LOCK) IN B5
A6 OUT LO-B & DF-LO T/H IN B6
A7
COM1 COM2
B7
A8 OUT HOLD END(H/E) SECTION IN B8
A9 OUT RUN D/Z IN B9
A10 OUT EVENT GRAPH TRIG IN B10
A11 OUT SIF CAL0 IN B11
A12 OUT SIF CAL1 IN B12
FCN-365P024-AU/FCN-360C024-BExtyernal I/O
- External I/O pin assignments for Comparison Standard B