EasyManuals Logo

YOKOGAWA DL850 ScopeCorder User Manual

YOKOGAWA DL850 ScopeCorder
190 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #49 background imageLoading...
Page #49 background image
Laser Specifications
Center wavelength: 850 nm
Pulse width: ≤ 10 ms (100 MHz), ≤ 2.5 ns (2 GHz)
Output: ≤ 1 mW
If the instrument is used in a manner not specified in this manual, the protection provided by the
instrument may be impaired. YOKOGAWA assumes no liability for the customer’s failure to comply
with these warnings and requirements.
LEDs on the Frequency Module
On the front panel of the frequency module (701280 (FREQ)) are LEDs for each channel. These LEDs
allow you to check the input condition of the pulse.
ACT
OVER
ACT
OVER
701280 (FREQ)
LED
Green: Pulses are being received.
Red: The input voltage range has
been exceeded.
Off: Pulses are not being received.
Note
The LEDs on the frequency module illuminates in green when pulse is applied and red when the input
voltage level is over range. It is independent of the start/stop condition of waveform acquisition of the
DL850/DL850V.
When the preset of the frequency module is set to EM Pickup (electromagnetic pickup), the LED will not
illuminate in red when the range is exceeded.
For information about presets, see section 1.7, “Configuring Frequency, Revolution, Period, Duty Cycle,
Power Supply Frequency, Pulse Width, Pulse Integration, and Velocity Measurements,” in the user’s
manual.
3.3 Installing Input Modules
3-8
IM DL850-03EN

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA DL850 ScopeCorder and is the answer not in the manual?

YOKOGAWA DL850 ScopeCorder Specifications

General IconGeneral
Maximum Sampling Rate100 MS/s
Power Supply100 to 240 VAC, 50/60 Hz
Resolution16-bit
Operating Temperature5 to 40°C
Bandwidth20 MHz
Input TypesVoltage, Temperature, Strain, Acceleration, Frequency, Logic, Current
InterfacesEthernet, USB, GP-IB
Trigger ModesEdge, Window, Pattern
Number of ChannelsUp to 128 (with multiple input modules)
Memory CapacityUp to 2 G points (with /M4 option)

Related product manuals