Maximum Input Voltages and Maximum Allowable Common Mode Voltages for the Modules
Applying a voltage exceeding the value indicated below may damage the input section. If the
frequency is above 1 kHz, damage may occur even when the voltage is below this value.
For the 720210 (HS100M12)
Maximum input voltage (at a frequency of 1 kHz or less)
When used with the 700929 (10:1) isolated probe or the 701947 (100:1) isolated probe.
1
1000 V (DC + ACpeak) CAT II
With the safety cable (1:1; a combination of the 701901 and 701954)
5
200 V (DC + ACpeak)
Direct input (cable that does not comply with the safety standards)
9
42V (DC + ACpeak)
Maximum allowable common mode voltage (at a frequency of 1 kHz or less)
When used with the 700929 (10:1) isolated probe or the 701947 (100:1) isolated probe.
2
Or when used with the safety cable (1:1; a combination of the 701901 and 701954).
8
1000 Vrms (CAT II)
Direct input (cable that does not comply with the safety standards)
10
42V (DC + ACpeak) (CAT II, 30 Vrms)
For the 701250 (HS10M12)
Maximum input voltage (at a frequency of 1 kHz or less)
When used with the 700929 (10:1) isolated probe or the 701947 (100:1) isolated probe.
1
600 V (DC + ACpeak)
Safety cable (1:1) (combined with 701901+701954)
5
200 V (DC + ACpeak) (as a value that meets the safety standard)
250 V (DC + ACpeak) ( maximum allowable voltage, as a value that does not damage
the instrument when applied)
Direct input (cable that does not comply with the safety standards
)
9
42V (DC + ACpeak)
Maximum allowable common mode voltage (at a frequency of 1 kHz or less)
When used with the 700929 (10:1) isolated probe or the 701947 (100:1) isolated probe.
2
Or when used with the safety cable (1:1; a combination of the 701901 and 701954).
8
400 Vrms (measurement category Other (O)), 300 Vrms (CAT II)
Direct input (cable that does not comply with the safety standards)
10
42 V (DC + ACpeak) (CAT II, 30 Vrms)
For the 701251 (HS1M16)
Maximum input voltage (at a frequency of 1 kHz or less)
When used with the 700929 (10:1) isolated probe or the 701947 (100:1) isolated probe.
1
600 V (DC + ACpeak)
Safety cable (1:1) (combined
with 701901 + 701954)
5
140 V (DC + ACpeak)
Direct input (cable that does not comply with the safety standards)
9
42V (DC + ACpeak)
Maximum allowable common mode voltage (at a frequency of 1 kHz or less)
When used with the 700929 (10:1) isolated probe or the 701947 (100:1) isolated probe.
2
Or when used with the safety cable (1:1; a combination of the 701901 and + 701954).
8
400 Vrms (measurement category Other (O)), 300 Vrms (CAT II)
Direct input (cable that does not comply with the safety standards)
10
42 V (DC + ACpeak) (CAT II, 30 Vrms)
For the 701255 (NONISO_10M12)
This module is non-isolated. Be sure to fasten the module screws when measuring a
voltage above 42 V on this module. In addition, use the 701940 (10:1) dedicated non-
isolated passive probe.
Maximum input voltage (at a frequency of 1 kHz or less)
Combined with the 701940 (10:1) passive probe
600 V (DC + ACpeak)
Di
rect input (cable that does not comply with the safety standards)
9
200 V (DC + ACpeak) (as a value that meets the safety standard)
250 V (DC + ACpeak) ( maximum allowable voltage, as a value that does not damage
the instrument when applied)
3.5 Connecting Probes
3-15
IM DL850-03EN
Making Preparations for Measurements
1
2
3
4
5
6
App
Index

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA DL850 ScopeCorder and is the answer not in the manual?

YOKOGAWA DL850 ScopeCorder Specifications

General IconGeneral
Maximum Sampling Rate100 MS/s
Power Supply100 to 240 VAC, 50/60 Hz
Resolution16-bit
Operating Temperature5 to 40°C
Bandwidth20 MHz
Input TypesVoltage, Temperature, Strain, Acceleration, Frequency, Logic, Current
InterfacesEthernet, USB, GP-IB
Trigger ModesEdge, Window, Pattern
Number of ChannelsUp to 128 (with multiple input modules)
Memory CapacityUp to 2 G points (with /M4 option)