TLOG Computation
In the explanation below, you cannot enter an expression that contains an operator or
function, an internal switch (S), a relay (I), or a flag (F) for e1. Also, an expression can
contain only one TLOG function.
TLOG.MAX( )
Maximum value
(Syntax) TLOG.MAX(e1)
(Condition) Determines the maximum value of channel e1.
TLOG.MIN( )
Minimum value
(Syntax) TLOG.MIN(e1)
(Condition) Determines the minimum value of channel e1.
TLOG.AVE( )
Average value
(Syntax) TLOG.AVE(e1)
(Condition) Determines the average value of channel e1.
TLOG.SUM( )
Sum value
(Syntax) TLOG.SUM(e1)
(Condition) Determines the sum of channel e1.
TLOG.P-P( )
Maximum – minimum value
(Syntax) TLOG.P-P(e1)
(Condition) Determines the maximum – minimum value of channel e1.
Expression Example
TLOG.MAX(01)+K01*SQR(02)
Examples of Invalid Expressions
TLOG.AVE(01)+TLOG.AVE(02)
Reason: TLOG appears twice in one expression.
TLOG.AVE(ABS(01))
Reason: A function is used inside the parentheses.
10.2 Writing Expressions

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA MVAdvanced MV1000 Series and is the answer not in the manual?

YOKOGAWA MVAdvanced MV1000 Series Specifications

General IconGeneral
BrandYOKOGAWA
ModelMVAdvanced MV1000 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals