Page #70 background image
3-6 IM 760301-01E
For Making Accurate Measurements
Allow the instrument to warm up for at least 30 minutes after turning ON the power
switch.
Perform zero-level compensation after warm-up (see section 10.1, “Performing Zero-
Level Compensation”).
Shutdown Operation
The setup parameters that exist immediately before the power switch is turned OFF are
stored in memory. The same is true when the power cord gets disconnected from the
outlet. The next time the power switch is turned ON, the instrument powers up using the
stored setup parameters.
Note
A lithium battery is used to retain the setup parameters. When the lithium battery voltage falls
below a certain level, a message is displayed on the screen (see section 11.2) when the
power switch is turned ON. If this message appears frequently, the battery must be replaced
quickly. The user cannot replace the battery. Contact your nearest YOKOGAWA dealer to
have the battery replaced. For information regarding battery life, see section 11.6.
3.4 Turning ON/OFF the Power Switch