5-19
IM 760301-01E
1
2
3
4
5
6
7
8
9
10
11
12
App
Index
Power Measurement
5.6 Measuring the Average Active Power
Explanation
The average active power is specified using the user-defined function as follows:
Average active power =
Integrated power
Elapsed time of integration
For example, to determine the average active power of element 1, the equation of the
user-defined function is set as follows:
WH(E1)/(TI(E1)/3600)