4-1
IM WT5000-02EN
4.1  Setting Integration Conditions
► “Enabling or Disabling Independent Integration (Independent Control)” in the features guide
► “Integration Auto Calibration On/Off (Auto Cal)” in the features guide
► “Integration Mode (Integration Mode)” in the features guide
► “Integration Timer (Integration Timer)” in the features guide
► “Scheduled Times for Real-Time Integration (Start Time/End Time)” in the features guide
► “Watt Hour Integration Method for Each Polarity (WP ± Type)” in the features guide
► “Current Mode for Current Integration (q Mode)” in the features guide
► “Rated Time for Integrated D/A Output (Integration Rated Time)” in the features guide
► “Integration resume action at power failure recovery (Resume Action)” in the features guide
This section explains operating procedures using the following setup methods.
•  Procedure Using the Setup Menu (see chapter 1)
•  Procedure Using the Menu Icons (see page iii)
•  Procedure Using the Keys (other than SETUP) (see section 1.2 in IM WT5000-03EN)
Procedure Using the Setup Menu
1.
  Tap the Setup icon  , or press MENU under SETUP.
2.
 Tap Computation/Output tab. A computation and output settings overview screen appears.
Pressing ESC closes the overview screen.
Setting Integration Conditions (Integration)
3.
 Tap Integration. An integration condition setup screen appears.
•  When the Integration Mode Is Set to Normal or Continuous
   
Computation/Output tab
Set the integration conditions.
Set the integration mode.
(Normal, Continuous, R-Normal, R-Continuous)
Set the integration timer.
Set the independent integration (OFF, ON).
Set integration auto calibration (OFF, ON).
Set the watt-hour integration 
methods for each polarity.
Set the current modes for 
current integration.
Set the integration resume operation at 
power failure recovery (Start, Stop, Error).
Chapter 4  Integrated Value Measurement (watt hours and ampere hours)