Section 10. Probe Configuration Download (ADFSET.EXE)
The
ADFSET
program allows you to download a probe configuration file to the ADVField when
operating from a PC-compatible computer using only a text interface (e.g., a palmtop computer).
This may be useful when changing probes in the field with limited computer access. When oper-
ating the ADV from the direct command interface (e.g.; using a data logger), it is important to
download the correct probe configuration file before starting data collection. If the wrong con-
figuration file is used, velocity data will be corrupted. The command line syntax is:
ADFSET [com port] [conf. file]
The
[com port]
parameter is the communication port being used on the computer (1 for COM1,
2 for COM2). The
[conf. file]
parameter is the name of the probe configuration file to be
downloaded to the ADV (e.g.,
1102.PRO
when using probe serial number 1102). The probe con-
figuration file must be located in the directory from which
ADFSET
is run.
After execution,
ADFSET
displays the series of commands sent to the ADV. It then shows the re-
sulting probe configuration as output by the ADV, starts ADV data collection (leaving the ADV
in data acquisition mode), and returns to the DOS prompt.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YSI SonTek ADVField and is the answer not in the manual?

YSI SonTek ADVField Specifications

General IconGeneral
BrandYSI
ModelSonTek ADVField
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals