SonTek/YSI
46
ADVField Operation Manual (September 1, 2001)
Show Deploy
• Displays the current autonomous deployment parameters.
• Example:
>show deploy
CURRENT DEPLOYMENT PARAMETERS
-----------------------------
Deployment - BA383
StartDate -- 2001/07/05
StartTime -- 17:44:10
-------------------------Sampling 1-----Sampling 2-----Sampling 3
SampRate (Hz) ----- 2.00 0.00 0.00
BurstInterval (s) - 1800 0 0
SamplesPerBurst --- 2048 0 0
RecordAmpCorr ----- YES YES YES
RecordCompass ----- NO NO NO
RecordSensor ------ YES YES YES
RecordExtSensor --- YES NO NO
RecordExtPress ---- NO NO NO
RecordStat -------- YES YES YES
RecordCtd --------- NO NO NO
RecordLisst-------- NO NO NO
Burst Length (s) -- 1024 0 0
Bytes per Burst --- 41058 0 0
Comments:
SonTek ADVField Autonomous System
We know how fast the water moves.
Do you?
3.12. Run Time Commands
The command shown here can be used while the ADVField is in data collection or deployment
mode. It is an alternative way to bring the instrument into command mode if it is not possible to
send a
BREAK
(e.g., if operating the system via a modem).
+++
(alternative
BREAK
command)
• Sending three + characters in succession causes the ADV to terminate data collection and re-
turn to the command mode.
• This command can be sent at any time during real-time data collection.
• In deployment mode, this command can only be used while the ADV is awake. It will not
work if the instrument is in the reduced power state between bursts.
• While the ADV is awake, there is no maximum time period during which the characters
must arrive.