ADVField/Hydra Operation Manual (September 1, 2001) 41
3.9. Sensor Commands
Sensor commands output data from the ADVField’s optional internal sensors: temperature, pres-
sure, and battery voltage. Additionally, a variety of optional external sensors can be controlled
through the direct command interface.
3.9.1. Internal Sensor Commands
Sensor [CONT]
Without parameter: Display once the current temperature (°C), pressure (dBar), and battery
voltage (VDC) data.
With
CONT
parameter: Display continuously (updated twice per second) the temperature
(°C), pressure (dBar), and battery voltage (VDC) data. Press any key to stop data output and
return to command mode.
Data is output in a self-explanatory, ASCII-text format.
Note: If
ExtSensor
=
3_OBS
, the pressure and battery voltage readings will display
N/A
.
PressScale
Displays the pressure-scaling factor used to convert counts to dBar. The scaling factor is
shown in units of dBar/count (see §4.8 for data conversion details).
PressOffset
Displays the pressure offset used to convert pressure counts to dBar. The offset is shown in
units of dBar (see §4.8 for data conversion details).
3.9.2. External Sensor Commands – Overall Control
These commands are common to all external sensors.
ExtSensor [NONE|STANDARD|3_OBS]
Without parameter: returns current setting.
When used with a valid parameter, tells the ADV processor what type of external sensor(s)
are installed.
NONE
– No external sensors are installed.
STANDARD
– One or two of the more typical external sensors is installed: SeaBird Micro-
Cat CTD (§3.9.3), a resonance pressure sensor (§3.9.4), a LISST sensor (§3.9.5).
3_OBS
– Three external sensors are installed. Note that with this option selected:
!
Sensor [CONT]
will display pressure and battery voltage readings as
N/A
.
! The real-time and deployment extraction software will store the Channel 3 sensor data
in the pressure column, but will not apply any pressure scaling to the data.
3.9.3. External Sensor Commands – SeaBird MicroCat CTD
H CTD
Displays help information for the CTD commands.
CTD [CONT]
Without parameter: Display once the most recent temperature (°C), conductivity (Siemens
per meter), pressure (decibar), and salinity (ppt) from the external CTD sensor.
With
CONT
parameter: Display continuously temperature (°C), conductivity (Siemens per
meter), pressure (decibar), and salinity (ppt) from the external CTD sensor. Press any key to
stop data output and return to command mode.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YSI SonTek ADVField and is the answer not in the manual?

YSI SonTek ADVField Specifications

General IconGeneral
BrandYSI
ModelSonTek ADVField
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals